Modified protocol for in vivo imaging of wild-type mouse retina with customized miniature spectral domain optical coherence tomography (SD-OCT) device

Abstract

This protocol outlines and evaluates a modified scanning procedure for a customized spectral domain optical coherence tomography (SD-OCT) imaging apparatus within the wild-type C57Bl/6 mouse posterior segment. This modified protocol allows for the capture of a 50 degree field of view spanning 3 mm by 3 mm perimeter with the optic disc as the central point. By utilizing this scanning protocol a more reliable measurement of retinal thickness can be achieved outside the fluctuating region of the optic disc. This protocol, when applied to this high resolution device, enables non-invasive in vivo histological imaging and biometric assessment of the various layers of the rodent posterior segment within a 20 – 30 min procedural time-frame. This protocol could establish a standardized method for evaluating morphological changes, with this commercial SDOCT device, when assessing longitudinal disease pathophysiology and treatment response in mouse models for future vision science research.

Background

Mouse models represent valuable paradigms for studying retinal neurodegenerative and vascular disorders that are analogous in humans. By characterizing the natural course of pathological change evident in these models, information pertaining to morphology and function can be ascertained for individual diseases. Mouse models have been instrumental for understanding pathophysiology in a variety of retinal diseases including: retinopathy of prematurity [1], diabetic retinopathy [24], exudative age-related macular degeneration [5], retinal vascular occlusion and ischemia-reperfusion injury models [6] and inherited retinal diseases [711].

The gold standard approach for studying retinal disease changes in these animal models has been through the use of ex vivo histological preparations after sacrificing the animal. This methodology is fraught with limitations as it provides one time observation and includes tissue damage, toxicity, laborious technical procedures, as well as tangential assessment of disease pathophysiology. In vivo techniques such as fundoscopy, confocal scanning laser ophthalmoscopy (cSLO), angiography, and electroretinography (ERG) allow longitudinal observation of dynamic functional and morphological changes within disease models, but do not provide histological evidence.

SD-OCT provides rapid, high resolution, cross-sectional images almost resembling histology images and assist in the diagnosis of posterior segment pathology at various time points [12, 13]. Because of this ability to acquire high-definition cross-sectional images of the retina, in humans, SD-OCT machines are being increasingly used in clinical practice. SD-OCT has been applied to investigate degenerative conditions not only in humans but also other animal species [1417]. SD-OCT provides histological – grade sections of the rodent posterior segment non-invasively. However, acquisition of images in animals is cumbersome as machines are specifically designed for human use.

In this report, we describe a modified scanning protocol for a stereotactic rotational multidirectional animal containment apparatus, with SD-OCT imaging capabilities, for the acquisition of high resolution scans of the ocular posterior segment in mice. Our objective was to evaluate the implementation of a modified scanning protocol on a commercially available miniature SD-OCT device customized for rodent retinal imaging. Current use of this SD-OCT apparatus relies on scanning protocols that isolate retinal areas only pertaining to scanning region of approximately 1.5 mm by 1.5 mm [18, 19]. Anatomically, this represents sections of greater thickness fluctuations as the retina begins to bottleneck with the emergence of the optic disc at the posterior pole. In order to measure a more uniform retinal region, measurements taken outside this central subfield (CSF) area would permit for a more accurate determination of retinal thickness. In our study we implemented a scanning protocol which delineated the CSF area in order to assess retinal thickness from more uniform regions of the posterior segment.

Results and discussion

The Bioptigen animal imaging mount with rodent alignment stage (AIM-RAS) apparatus and miniature SD-OCT hand-held probe (HHP), allows for effective visualization of the posterior segment in wild-type C57Bl/6 mice. Figures 1A and 1B demonstrate en-face as well as B-scan images of a representative posterior segment scan from a C57Bl/6 wild-type mouse while utilizing the modified scanning protocol for the HHP. As depicted in Figure 1C, the retinal layers which were measured included the following: choriocapillaris (CC), outer retinal layer (ORL), inner retinal layer (IRL), retinal pigment epithelium (RPE), external limiting membrane/inner segment of photoreceptors/outer segment of photoreceptors (ELM/IS/OS), outer nuclear layer (ONL), outer plexiform layer (OPL), inner nuclear layer (INL), inner plexiform layer /ganglion cell (IPL/GC), and retinal nerve fiber layer (RNFL). Figure 2 graphically depicts the thickness measurements for three mice imaged with the AIM-RAS apparatus. Overall, the mean (range) value for the thickness measurements for the CC, ORL, and IRL were 42.67 μm (39 μm – 48 μm), 103 μm (97 μm – 111 μm), and 120 μm (118 μm – 124 μm) respectively. Measuring each layer of the retina individually, the mean (range) thickness values for the RPE was 16.67 μm (13 μm – 20 μm); ELM/IS/OS was 23.33 μm (17 μm – 29 μm); ONL was 63 μm (60 μm – 65 μm); OPL was 15.67 μm (14 μm – 17 μm), INL was 30 μm (29 μm – 32 μm), IPL/GC was 56.33 μm (52 μm – 61 μm), and RNFL was 18 μm (14 μm – 21 μm). The mean thickness magnitudes for the measured layers were not significantly different among the animals studied, F (2, 27) = 0.022 (p = 0.978).

Figure 1
figure 1

C57Bl/6 retinal scans and posterior segmental layer thickness analysis. En-face view of posterior segment with centralized optic nerve head (a). B-scan image of mouse retina from cross-sectional scan of en-face image (b). B-scan image with defined regions and thickness of posterior segment layers (c). CC: choriocapillaris, ORL: outer retinal layer, IRL: inner retinal layer, RPE: retinal pigment epithelium, ELM/IS/OS: external limiting membrane/inner segment of photoreceptors/outer segment of photoreceptors, ONL: outer nuclear layer, OPL: outer plexiform layer, INL: inner nuclear layer, IPL/GC: inner plexiform layer /ganglion cell, RNFL: retinal nerve fiber layer.

Figure 2
figure 2

Thickness measurements of C57Bl/6 posterior segment layers. Thickness was calculated utilizing InVivoVue Clinic imaging software. Abbreviations represent the following: CC: choriocapillaris, ORL: outer retinal layer, RPE: retinal pigment epithelium, ELM/IS/OS: external limiting membrane/inner segment of photoreceptors/outer segment of photoreceptors, ONL: outer nuclear layer, IRL: inner retinal layer, OPL: outer plexiform layer, INL: inner nuclear layer, OPL: outer plexiform layer, IPL/GC: inner plexiform layer /ganglion cell, RNFL: retinal nerve fiber layer.

The Bioptigen spectral domain ophthalmic imaging system (SDOIS) represents an alternative non-invasive in vivo imaging device which achieves high resolution histological – grade cross-sections along with en-face retinal imaging for the purposes of amassing volume intensity projection scans. The SDOIS possesses comparable axial resolution depths as observed with other high resolution OCT devices [14, 15, 1821]. The use of this specific setup, along with the customized scanning protocol designed to capture full-field fundus images and cross-sectional views of the posterior segment, permits the acquisition of real-time SD-OCT scans of retinal morphology.

Conclusion

In summary, this protocol details the animal and ophthalmic preparatory procedures as well as optimized technical method for acquiring retinal scans with the SDOIS. The imaging procedures as outlined in this protocol can be employed to assess normal mouse retinal anatomy along with rodent ocular disease model morphology when operating the SDOIS AIM-RAS setup.

Methods

Materials and supplies

Three adult C57Bl/6 mice (Jackson Laboratory, Bar Harbor, ME) were utilized for this study. All experiments were performed in accordance with the University of Florida institutional animal care and use committee guidelines and adhered to The Code of Ethics of the World Medical Association (Declaration of Helsinki) for animal experiments. The mice were housed and maintained in the University of Florida animal care services (ACS) facility and subjected to a 12-hour light/dark cycle with constant access to nourishments. All scanning and preparatory procedures were conducted in an ACS designated and approved procedural location.

Animal preparation

Ketamine (Ketaject, 100 mg/mL, 80 mg/kg body weight) and Xylazine (Ana Sed, 100 mg/mL, 10 mg/kg body weight) anesthetics were purchased from Webster Veterinary (Devens, MA). Sterile NaCl 0.9% (wt/vol) was used in addition to Ketamine and Xylazine for creation of the anesthetic mixture. The Ketamine (16 mg/mL) and Xylazine (2 mg/mL) anesthesia stock reagents were diluted in sterile 0.9% NaCl solution. Mice were anesthetized by means of i.p. injection of Ketamine/Xylazine mixture. Afterwards, the animals were placed onto a 37°C thermostatic heating pad while in a sternally recumbent position. Once the mice were deemed fully anesthetized, one drop of dilating agent was applied for 2 – 3 minutes bilaterally. The dilating solution was wicked away and one drop of artificial tear was added to each eye to prevent corneal desiccation. Mydriasis was achieved with use of Tropicamide 1% (Akorn Inc.; Lake Forest, IL) while corneal hydration was maintained with Systane Ultra® lubricant eye drops (Alcon, Fort Worth, TX).

Imaging and animal equipment

The Bioptigen SDOIS (Bioptigen, Inc., Durham, NC) is a noninvasive imaging Class I, Type B, IPXO, continuous operation medical device. It delivers axial resolution capabilities of less than 6 microns and can process 20,000 A-scans/second. The SDOIS apparatus is comprised of a base system as well as an animal imaging mount and rodent alignment stage (AIM-RAS), which houses a SD-OCT hand held probe (HHP) (Figure 3a – c). The base system incorporates a host computer, 840 nm OCT engine with reference arm attachment, and the HHP. The HHP SD-OCT scanner was encased via the Animal Imaging Mount (AIM) (Bioptigen, Inc., Durham, NC), which allowed forward and backward adjustment of the HHP (Figure 3b). The InVivoVue Clinic software (Bioptigen, Inc., Durham, NC) enables the creation, display, loading and saving of OCT image files. The Rodent Alignment System (RAS) contained an X- (micrometer), Y- (scissor jack), and Z- translators along with stereotactic rotational cassette (for housing mice) within a bushing and platform base (Bioptigen, Inc., Durham, NC) (Figure 3c). The entire AIM-RAS device was attach to a slit-lamp base. The aiming tip, with cassette tongue, and aiming target were used for determining geometrical working distance for the optical lens bore to the mouse eye (Bioptigen, Inc., Durham, NC) (Figure 4a). In order to suppress breathing artifacts and provide fine tuning of horizontal and vertical positioning, a bite bar was employed (Bioptigen, Inc., Durham, NC) (Figure 4a). Cotton-tip applicators (Rush Dental & Medical Supply; Santa Monica, CA) were employed to wick eye drops (Figure 4b). In an attempt to wedge the mouse to a selective side of the animal cassette, as well as blanket the animal to maintain body warmth, surgical gauze was utilized (AllegroMedical) (Figure 4b). Surgical (3M™) and marking (Sigma) tape were applied so as to secure the mouse during rotational maneuvering and to identify the location of Z-translator when aligning the aiming tip to aiming target (Figure 4b).

Figure 3
figure 3

Imaging setup. Overview of the Bioptigen spectral domain ophthalmic imaging system (SDOIS) (a), Animal Imaging Mount (AIM) component (b), and Rodent Alignment System (RAS) component (c). AIM component includes miniature SD-OCT hand-held probe (1) and Z-translator (2). RAS component includes cassette (3), bushing (4), base (5), X-translator (6), and Y-translator (7).

Figure 4
figure 4

Accessories for animal imaging mount with rodent alignment stage (AIM-RAS) setup. Aiming target with tongue (arrow) and mouse bite bar (a). Marking tape, mouse bedding/scaffold, rolled up gauze for wedging mouse, and cotton-tip applicator for wicking eye drops (b).

Mouse positioning

The mice were positioned into the animal cassette according to body habitus. Depending on mouse body size, smaller mice were raised within the cassette by fashioning rectangular bedding made from gauze or folded paper towel. This assured that the mouse was positioned within a straight line and that the neck was level so that the head was looking straight forward. Also, depending on which eye was selected for scanning, a cylindrical gauze roll was created in order to bias the mouse body, and therefore eye, to one side of the cassette. Thus, when scanning the right eye, the animal was biased toward the left side of the cassette (Figure 5). Lubricant drops were evenly spread across the cornea during rewetting in order to eliminate optical aberrancies and ensure adequate focus during scanning of the retina. The mouse was secured within the cassette via comfortably strapping the animal with surgical tape. The HHP bore, with aiming target attached, was brought in close proximity to eye by adjusting the Z-translator on the AIM apparatus or by shifting the cassette within the bushing.

Figure 5
figure 5

Mouse insertion into cassette and attachment to bite bar. Aerial view of mouse within cassette while attached to bite bar; large white arrow points to right eye where all scans were obtained from; small white arrow indicates gauze roll used for wedging animal to one side of the cassette; black arrow shows bedding/scaffold used to maintain straight and level head and body posture for the mouse (a). Lateral view of mouse within cassette (b). Lateral view of mouse blanketed with gauze to retain body heat (c).

SD-OCT imaging of ONH and retina

After the HHP lens was situated close to the right eye of the animal, the InVivoVue Clinic application was activated and the scanning began- following setup of subject profile for image acquisition. We selected the rectangular scanning protocol consisting of a 3 mm by 3 mm perimeter with 1000 A-scans per B-scan with a total B-scan amount of 100. This is a modification of the recommended parameters of 1.4 mm by 1.4 mm set by the company for performing rectangular scans.

In order to bring the ONH into view and focus we made fine tune adjustments to the animal cassette. At this point the eye was docked into the aiming target, and the ONH was centered within the viewing pane, of the InVivoVue Clinic application, for the B-scan and the en face images of the retina. We fine-tuned B-scan focusing by adjusting the Z-translator on the AIM, which moved the HHP towards or away from the eye, and also by dialing the reference arm on the base system in order to obtain bright crisp B-scan images. The CSF was defined as the central 1 mm area containing the ONH. This region was demarcated by using the InvVivoVue Clinic measurement calipers to establish the 1 mm diameter area from the center of the ONH. Thickness measurements were performed with the caliper instruments outside the CSF area.

Statistical analysis

Statistical analyses included the mean and range for each of the posterior segment layers in the three animals examined. A one-way analysis of variance was performed to contrast layer thickness between the three animals. Statistical significance was achieved if the two tailed p-value was less than the alpha level of 0.05.

References

  1. Smith LE, Wesolowski E, McLellan A, Kostyk SK, D'Amato R, Sullivan R, D'Amore PA: Oxygen-induced retinopathy in the mouse. Invest Ophthalmol Vis Sci. 1994, 35 (1): 101-111.

    CAS  PubMed  Google Scholar 

  2. Kern TS, Engerman RL: A mouse model of diabetic retinopathy. Arch Ophthalmol. 1996, 114 (8): 986-990. 10.1001/archopht.1996.01100140194013.

    Article  CAS  PubMed  Google Scholar 

  3. Feit-Leichman RA, Kinouchi R, Takeda M, Fan Z, Mohr S, Kern TS, Chen DF: Vascular damage in a mouse model of diabetic retinopathy: relation to neuronal and glial changes. Invest Ophthalmol Vis Sci. 2005, 46 (11): 4281-4287. 10.1167/iovs.04-1361.

    Article  PubMed  Google Scholar 

  4. Zheng L, Du Y, Miller C, Gubitosi-Klug RA, Ball S, Berkowitz BA, Kern TS: Critical role of inducible nitric oxide synthase in degeneration of retinal capillaries in mice with streptozotocin-induced diabetes. Diabetologia. 2007, 50 (9): 1987-1996. 10.1007/s00125-007-0734-9.

    Article  CAS  PubMed  Google Scholar 

  5. Aguilar E, Dorrell MI, Friedlander D, Jacobson RA, Johnson A, Marchetti V, Moreno SK, Ritter MR, Friedlander M: Chapter 6. Ocular models of angiogenesis. Methods Enzymol Volume 444. Edited by: Cheresh DA. 2008, New York: Elsevier Inc, 115-158.

    Google Scholar 

  6. Grant MB, May WS, Caballero S, Brown GA, Guthrie SM, Mames RN, Byrne BJ, Vaught T, Spoerri PE, Peck AB, Scott EW: Adult hematopoietic stem cells provide functional hemangioblast activity during retinal neovascularization. Nat Med. 2002, 8 (6): 607-612. 10.1038/nm0602-607.

    Article  CAS  PubMed  Google Scholar 

  7. Karan G, Lillo C, Yang Z, Cameron DJ, Locke KG, Zhao Y, Thirumalaichary S, Li C, Birch DG, Vollmer-Snarr HR, Williams DS, Zhang K: Lipofuscin accumulation, abnormal electrophysiology, and photoreceptor degeneration in mutant ELOVL4 transgenic mice: a model for macular degeneration. Proc Natl Acad Sci U S A. 2005, 102 (11): 4164-4169. 10.1073/pnas.0407698102.

    Article  PubMed Central  CAS  PubMed  Google Scholar 

  8. Pang JJ, Chang B, Hawes NL, Hurd RE, Davisson MT, Li J, Noorwez SM, Malhotra R, McDowell JH, Kaushal S, Hauswirth WW, Nusinowitz S, Thompson DA, Heckenlively JR: Retinal degeneration 12 (rd12): a new, spontaneously arising mouse model for human Leber congenital amaurosis (LCA). Mol Vis. 2005, 11: 152-162.

    CAS  PubMed  Google Scholar 

  9. Pittler SJ, Baehr W: Identification of a nonsense mutation in the rod photoreceptor cGMP phosphodiesterase beta-subunit gene of the rd mouse. Proc Natl Acad Sci U S A. 1991, 88 (19): 8322-8326. 10.1073/pnas.88.19.8322.

    Article  PubMed Central  CAS  PubMed  Google Scholar 

  10. Radu RA, Mata NL, Bagla A, Travis GH: Light exposure stimulates formation of A2E oxiranes in a mouse model of Stargardt's macular degeneration. Proc Natl Acad Sci U S A. 2004, 101 (16): 5928-5933. 10.1073/pnas.0308302101.

    Article  PubMed Central  CAS  PubMed  Google Scholar 

  11. Weber BH, Schrewe H, Molday LL, Gehrig A, White KL, Seeliger MW, Jaissle GB, Friedburg C, Tamm E, Molday RS: Inactivation of the murine X-linked juvenile retinoschisis gene, Rs1h, suggests a role of retinoschisin in retinal cell layer organization and synaptic structure. Proc Natl Acad Sci U S A. 2002, 99 (9): 6222-6227. 10.1073/pnas.092528599.

    Article  PubMed Central  CAS  PubMed  Google Scholar 

  12. Alam S, Zawadzki RJ, Choi S, Gerth C, Park SS, Morse L, Werner JS: Clinical application of rapid serial fourier-domain optical coherence tomography for macular imaging. Ophthalmology. 2006, 113 (8): 1425-1431. 10.1016/j.ophtha.2006.03.020.

    Article  PubMed Central  PubMed  Google Scholar 

  13. Wojtkowski M, Srinivasan V, Fujimoto JG, Ko T, Schuman JS, Kowalczyk A, Duker JS: Three-dimensional retinal imaging with high-speed ultrahigh-resolution optical coherence tomography. Ophthalmology. 2005, 112 (10): 1734-1746. 10.1016/j.ophtha.2005.05.023.

    Article  PubMed Central  PubMed  Google Scholar 

  14. Fischer MD, Huber G, Beck SC, Tanimoto N, Muehlfriedel R, Fahl E, Grimm C, Wenzel A, Remé CE, van de Pavert SA, Wijnholds J, Pacal M, Bremner R, Seeliger MW: Noninvasive, in vivo assessment of mouse retinal structure using optical coherence tomography. PLoS One. 2009, 4 (10): e7507-10.1371/journal.pone.0007507.

    Article  PubMed Central  PubMed  Google Scholar 

  15. Huber G, Beck SC, Grimm C, Sahaboglu-Tekgoz A, Paquet-Durand F, Wenzel A, Humphries P, Redmond TM, Seeliger MW, Fischer MD: Spectral domain optical coherence tomography in mouse models of retinal degeneration. Invest Ophthalmol Vis Sci. 2009, 50 (12): 5888-5895. 10.1167/iovs.09-3724.

    Article  PubMed Central  PubMed  Google Scholar 

  16. Srinivasan VJ, Ko TH, Wojtkowski M, Carvalho M, Clermont A, Bursell SE, Song QH, Lem J, Duker JS, Schuman JS, Fujimoto JG: Noninvasive volumetric imaging and morphometry of the rodent retina with high-speed, ultrahigh-resolution optical coherence tomography. Invest Ophthalmol Vis Sci. 2006, 47 (12): 5522-5528. 10.1167/iovs.06-0195.

    Article  PubMed Central  PubMed  Google Scholar 

  17. Wolf-Schnurrbusch UE, Enzmann V, Brinkmann CK, Wolf S: Morphologic changes in patients with geographic atrophy assessed with a novel spectral OCT-SLO combination. Invest Ophthalmol Vis Sci. 2008, 49 (7): 3095-3099. 10.1167/iovs.07-1460.

    Article  PubMed  Google Scholar 

  18. Pennesi ME, Michaels KV, Magee SS, Maricle A, Davin SP, Garg AK, Gale MJ, Tu DC, Wen Y, Erker LR, Francis PJ: Long-Term Characterization of Retinal Degeneration in rd1 and rd10 Mice Using Spectral Domain Optical Coherence Tomography. Invest Ophthalmol Vis Sci. 2012, 10;53 (8): 4644-4656.

    Article  Google Scholar 

  19. Kagemann L, Wollstein G, Ishikawa H, Nadler Z, Sigal IA, Folio LS, Schuman JS: Visualization of the conventional outflow pathway in the living human eye. Ophthalmology. 2012, 119 (8): 1563-1568. 10.1016/j.ophtha.2012.02.032.

    Article  PubMed Central  PubMed  Google Scholar 

  20. Fukuchi T, Takahashi K, Shou K, Matsumura M: Optical coherence tomography (OCT) findings in normal retina and laser-induced choroidal neovascularization in rats. Graefes Arch Clin Exp Ophthalmol. 2001, 239 (1): 41-46. 10.1007/s004170000205. Erratum in: Graefes Arch Clin Exp Ophthalmol 2001, 239(5):387

    Article  CAS  PubMed  Google Scholar 

  21. Ruggeri M, Wehbe H, Jiao S, Gregori G, Jockovich ME, Hackam A, Duan Y, Puliafito CA: In vivo three-dimensional high-resolution imaging of rodent retina with spectral-domain optical coherence tomography. Invest Ophthalmol Vis Sci. 2007, 48 (4): 1808-1814. 10.1167/iovs.06-0815.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

We like to thank James M. Dominguez, Lynn Shaw, and Maria Grant for providing access to animal subjects and use of their laboratory facilities.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Kakarla V Chalam.

Additional information

Competing interests

The authors declare that they have no competing financial interests.

Authors’ contributions

L.R.F. and S.B. prepped the animals, performed scans, and optimized the procedure. L.R.F. wrote the manuscript, collected the data, presented and prepared the figures. S.G. and K.V.C. assisted in structuring the content of the manuscript and revisions. K.V.C. supervised the work. All authors read and approved the final manuscript.

Authors’ original submitted files for images

Rights and permissions

Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Ferguson, L.R., Balaiya, S., Grover, S. et al. Modified protocol for in vivo imaging of wild-type mouse retina with customized miniature spectral domain optical coherence tomography (SD-OCT) device. Biol Proced Online 14, 9 (2012). https://doi.org/10.1186/1480-9222-14-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1480-9222-14-9

Keywords